Correlation between plasma renin concentration and plasma renin activity assays on the Maglumi 800 chemiluminescence immunoassay analyzer
Tác giả: Luong Hue Quyen, Tran Thi Chi MaiTóm tắt:
Evaluated the correlation between plasma renin concentration (PRC) and plasma renin activity (PRA), and then PAC/PRC and PAC/PRA ratios on Maglumi 800 immunoassay analyzer. Forty-five plasma samples from volunteers were collected. PRC, PRA and PAC were measured by Maglumi 800 - a fully-auto chemiluminescence immunoassay analyzer. Correlation between plasma renin concentration (PRC) and plasma renin activity (PRA), and then PAC/PRC and PAC/PRA ratios were determined by regression analysis. PRC and PRA showed a good correlation (r = 0.88). The PAC/PRC and PAC/PRA ratios have a similar relationship (r = 0.757). Because of the advantages of the procedure and the independence of PRC from endogenous angiotensinogen levels, PRC may be used preferentially over PRA as high throughput screening tool.